SW-2 SWAB PROCESSING
REDI-NET Consortium
Disclaimer
This work is supported by the US Army Medical Research and Development Command under Contract No.W81XWH-21-C-0001, W81XWH-22-C-0093 and HT9425-23-C-0059. The views, opinions and/or findings contained in this report are those of the author(s) and should not be construed as an official Department of the Army or Navy position, policy or decision unless so designated by other documentation.
Abstract
This protocol details procedures for total nucleic acid extraction from swab samples.
Before start
BEFORE START
- Pre-cool the Bullet Blender by adding dry ice into the cooling compartment and running the cooling program.
- Clean the work surfaces with RNaseZap, then wipe the surfaces with 70% molecular biology grade ethanol to remove additional contaminants.
- Pre-heat the heater mixer at
56°C
. - Transfer 0.1 mm zirconium oxide beads (2 spoons, Appendix 1) to Thermo Scientific Screw Cap Micro 1.5ml Tubes.
- Buffer ATL may form precipitates upon storage. If necessary, warm to
56°C
until the precipitates have fully dissolved. Prepare buffer ATL-DX: add100µL
Reagent DX to15mL
Buffer ATL. If smaller amounts are needed, transfer1.5mL
of Buffer ATL into a sterile 2ml vial and add10µL
Reagent DX. Mix well, after addition of Reagent DX. After preparation, the mixture is stable for 6 months atRoom temperature
(15-25°C). - For the first time use of IndiMag pathogen kit, add 100% ethanol to Buffer AW1 and AW2, and add 100% isopropanol to ACB as indicated on the bottles.
- MagAttract Suspension G from IndiMag pathogen kit needs to be vortexed thoroughly for
0h 3m 0s
(before first use) or 1 minute (before subsequent uses) to ensure that the magnetic silica particles are fully resuspended. - Aliquot nuclease-free water in big bottle into a few 15ml tubes for preparing TNA elution in KingFisher Flex or KingFisher Duo Prime to avoid cross-contamination.
Steps
1. SAMPLE LYSIS
500µL
of ATL-DX buffer and 20µL
Proteinase K to the bead tubes prepared in Step 4 of Before Start section under the Guidelines & Warning tab. Place the sample swab in the bead tube with ATL buffer and Proteinase K. Use sterile scissors to carefully cut off the swab heads or transfer the pre-cut swab head from the original container to the bead tube with forceps.
Include a positive control for each batch of samples: transfer 37.5µL
ZymoBIOMICS Microbial Community Standard Material, 100µL
EBV, and 100µL
ZIKV standard into a tube from Step 3 of Before Start section. Add 250µL
ATL-DX buffer and 20µL
proteinase K.
Include a negative control for each batch of samples: a bead tube from Step 3 of Before Start section with 500µL
ATL-DX buffer and 20µL
proteinase K.
Incubate samples in a thermomixer at 1400rpm
.
Refill the dry ice compartment of the Bullet Blender if necessary. After incubation, load all samples into the Bullet Blender.
Set the speed at 12 and the time at 3. Press Start.
Let the samples settle for 0h 1m 0s
in the Bullet Blender and then repeat Step 7.
2. INSTRUMENT SET UP
Confirm 96 deep-well magnetic heads and 96 well deep-well heat blocks are being used.
Ensure the program IndiMag_Pathogen_KF_Flex_4wash or the program has been downloaded and loaded onto the KingFisher Flex instrument.
2.1 SET UP THE PROCESSING PLATES
Set up the Tip Comb, Wash, and Elution Plates outside the instrument according to the following table.
A | B | C | D | E |
---|---|---|---|---|
Plate ID | Plate position | Plate type | Reagent | Volume per well |
Tip comb | 7 | Place a 96 Deep-well Tip comb in a deep-well plate | ||
Elution | 6 | Deep-Well | Nuclease-free water | 75 µL |
Wash 4 | 5 | Deep-Well | 100% ethanol | 750 µL |
Wash 3 | 4 | Deep-Well | 80% ethanol | 750 µL |
Wash 2 | 3 | Deep-Well | Buffer AW2 | 700 µL |
Wash 1 | 2 | Deep-Well | Buffer AW1 | 700 µL |
Sample | 1 | Sample Lysate | Lysate and lysis buffer | 985 µL |
2.2 EXTRACTION
Centrifuge the bead tubes with lysate from Step 8 for 12000x g
.
Transfer 290µL
supernatant without any particle carryover to the wells of the Deep-well plate. This plate becomes the Sample Plate.
Add 135µL
Buffer VXL, 540µL
Buffer ACB, and 20µL
MagAttract Suspension G to each sample in the sample plate. For multiple samples, make a master mix with 10% overage. Invert slowly to mix the master mix; avoid foaming (it can be mixed on a Hula mixer for 2 min). Add 695µL
mixture to each sample. Each well including sample lysate should be 985µL
.
Select the program IndiMag_Pathogen_KF_Flex_4wash on the instrument.
Start the run, then load the prepared plates into position when prompted by the instrument.
2.3 QUANTIFICATION AND STORAGE
After the running protocol is completed (~35 minutes), immediately remove the elution plate from the instrument and cover the plate or transfer the eluate to the final tube or plate of choice for final storage.
In a 0.6mL microcentrifuge tube, use 3µL
total nucleic acid for DNA and RNA concentration measurement using Qubit 4 Fluorometer following manufacturer instructions (Kits needed: Qubit 1X dsDNA HS Assay Kit and Qubit RNA HS Assay Kit) (see Appendix 2 and Appendix 3).
Proceed with sample testing following the REDI-NET SOP SW-4 Swab Testing or store at -20°C
for less than 2 weeks (for long-term storage the sample needs to be stored at -80°C
following the REDI-NET SOP SW-3 Swab Storage).
3. INSTRUMENT SET UP
Confirm 12-tip magnetic heads and 12 well deep-well heat blocks are being used.
Ensure the program IndiMag_Pathogen_KF_Duo_4wash has been downloaded and loaded onto the KingFisher Duo Prime instrument.
3.1 SET UP THE SAMPLE PLATE AND ELUTION STRIP
Set up the Sample Plate according to the table below:
A | B | C | D |
---|---|---|---|
Row ID | Plate Row | Reagent | Volume per well |
Sample row | A | Lysate and lysis buffer | 985 µL |
Wash 1 | B | Buffer AW1 | 700 µL |
Wash 2 | C | Buffer AW2 | 700 µL |
Wash 3 | D | 80% ethanol | 750 µL |
Wash 4 | E | 100% ethanol | 750 µL |
Tip Comb | F | Tip comb | |
**** | G | Empty | |
**** | H |
Set up the Elution Strip according to the table below:
A | B | C | D |
---|---|---|---|
Strip ID | Row | Reagent | Volume per well |
Elution | A | Nuclease-free water | 75 µL |
3.2 EXTRACTION
Centrifuge the bead tubes with lysate from Step 8 for 12000x g
.
Transfer 290µL
supernatant without any particle carryover to the wells of the Deep-well Row A. This row becomes the Sample Row.
Add 135µL
Buffer VXL, 540µL
Buffer ACB, and 20µL
MagAttract Suspension G to each sample in the Sample Plate. For multiple samples, make a master mix with 10% overage. Invert slowly to mix the master mix and avoid foaming. Add 695µL
mixture to each sample. Each well including sample lysate should be 985µL
.
Select the program IndiMag_Pathogen_KF_Duo_4wash on the instrument.
Start the run, then load the prepared plate and Elution Strip into position when prompted by the instrument.
3.3 QUANTIFICATION AND STORAGE
After the running protocol is completed (~35 minutes), immediately remove the elution plate from the instrument and cover the plate or transfer the eluate to the final tube or plate of choice for final storage.
In a 0.6mL microcentrifuge tube, use 3µL
total nucleic acid for DNA and RNA concentration measurement using Qubit 4 Fluorometer following manufacturer instructions (Kits needed: Qubit 1X dsDNA HS Assay Kit and Qubit RNA HS Assay Kit) (see Appendix 2 and Appendix 3).
Proceed with sample testing following the REDI-NET SOP SW-4 Swab Testing or store at -20°C
for less than 2 weeks (for long-term storage the sample needs to be stored at -80°C
following the REDI-NET SOP SW-3 Swab Storage).
APPENDIX 2. DNA and RNA Measurement using QUBIT FLUOROMETER 4.0
DNA quantification:
According to the volume of sample used, add the 1x HS dsDNA Qubit Assay for a final volume of 200µL
(i.e., if using 3µL
of sample, add 197µL
of 1x HS dsDNA Qubit Assay). Vortex for 5 - 10 seconds, then incubate for 0h 2m 0s
at Room temperature
RNA Quantification:
In a new microcentrifuge tube/falcon tube (depending on the number of samples processed), prepare a working solution of the Qubit HS RNA Assay:
A | B | C |
---|---|---|
Reagents | Volume/sample | Volume for n+1 sample |
Qubit RNA HS Assay buffer | 199 µL | …. µL |
Qubit RNA HS Assay Dye | 1 µL | …. µL |
In a new 0.6ml tube, mix 197µL
of Qubit HS RNA Assay working solution and 3µL
of the sample. Vortex for 5 - 10 seconds, then incubate for 0h 2m 0s
at Room temperature
before reading.