SW-2 SWAB PROCESSING

REDI-NET Consortium

Published: 2024-02-02 DOI: 10.17504/protocols.io.eq2lyjjkrlx9/v1

Disclaimer

This work is supported by the US Army Medical Research and Development Command under Contract No.W81XWH-21-C-0001, W81XWH-22-C-0093 and HT9425-23-C-0059. The views, opinions and/or findings contained in this report are those of the author(s) and should not be construed as an official Department of the Army or Navy position, policy or decision unless so designated by other documentation.

Abstract

This protocol details procedures for total nucleic acid extraction from swab samples.

Before start

BEFORE START

  1. Pre-cool the Bullet Blender by adding dry ice into the cooling compartment and running the cooling program.
  2. Clean the work surfaces with RNaseZap, then wipe the surfaces with 70% molecular biology grade ethanol to remove additional contaminants.
  3. Pre-heat the heater mixer at 56°C.
  4. Transfer 0.1 mm zirconium oxide beads (2 spoons, Appendix 1) to Thermo Scientific Screw Cap Micro 1.5ml Tubes.
  5. Buffer ATL may form precipitates upon storage. If necessary, warm to 56°C until the precipitates have fully dissolved. Prepare buffer ATL-DX: add 100µL Reagent DX to 15mL Buffer ATL. If smaller amounts are needed, transfer 1.5mL of Buffer ATL into a sterile 2ml vial and add 10µL Reagent DX. Mix well, after addition of Reagent DX. After preparation, the mixture is stable for 6 months at Room temperature (15-25°C).
  6. For the first time use of IndiMag pathogen kit, add 100% ethanol to Buffer AW1 and AW2, and add 100% isopropanol to ACB as indicated on the bottles.
  7. MagAttract Suspension G from IndiMag pathogen kit needs to be vortexed thoroughly for 0h 3m 0s (before first use) or 1 minute (before subsequent uses) to ensure that the magnetic silica particles are fully resuspended.
  8. Aliquot nuclease-free water in big bottle into a few 15ml tubes for preparing TNA elution in KingFisher Flex or KingFisher Duo Prime to avoid cross-contamination.

Steps

1. SAMPLE LYSIS

1.

Note
NOTE: To prevent contamination samples nucleic acid extraction and amplification (PCR) should be performed in separate rooms.
Add 500µL of ATL-DX buffer and 20µL Proteinase K to the bead tubes prepared in Step 4 of Before Start section under the Guidelines & Warning tab.

2.

Place the sample swab in the bead tube with ATL buffer and Proteinase K. Use sterile scissors to carefully cut off the swab heads or transfer the pre-cut swab head from the original container to the bead tube with forceps.

3.

Include a positive control for each batch of samples: transfer 37.5µL ZymoBIOMICS Microbial Community Standard Material, 100µL EBV, and 100µL ZIKV standard into a tube from Step 3 of Before Start section. Add 250µL ATL-DX buffer and 20µL proteinase K.

4.

Include a negative control for each batch of samples: a bead tube from Step 3 of Before Start section with 500µL ATL-DX buffer and 20µL proteinase K.

5.

Incubate samples in a thermomixer at 1400rpm.

6.

Refill the dry ice compartment of the Bullet Blender if necessary. After incubation, load all samples into the Bullet Blender.

7.

Set the speed at 12 and the time at 3. Press Start.

8.

Let the samples settle for 0h 1m 0s in the Bullet Blender and then repeat Step 7.

Note
STOPPING POINT : lysed samples can be stored at 4°C 0h 1m 0s.

2. INSTRUMENT SET UP

9.

Note
NOTE: KingFisher Flex only, if using KingFisher Duo Prime, go to section 3

Confirm 96 deep-well magnetic heads and 96 well deep-well heat blocks are being used.

10.

Ensure the program IndiMag_Pathogen_KF_Flex_4wash or the program has been downloaded and loaded onto the KingFisher Flex instrument.

2.1 SET UP THE PROCESSING PLATES

11.

Set up the Tip Comb, Wash, and Elution Plates outside the instrument according to the following table.

Note
NOTE: DO NOT use the elution buffer provided by the kit for TNA elution. The ingredients in the elution buffer inhibit the downstream DNA sequencing efficiency.

ABCDE
Plate IDPlate positionPlate typeReagentVolume per well
Tip comb7Place a 96 Deep-well Tip comb in a deep-well plate
Elution6Deep-WellNuclease-free water75 µL
Wash 45Deep-Well100% ethanol750 µL
Wash 34Deep-Well80% ethanol750 µL
Wash 23Deep-WellBuffer AW2700 µL
Wash 12Deep-WellBuffer AW1700 µL
Sample1Sample LysateLysate and lysis buffer985 µL

2.2 EXTRACTION

12.

Centrifuge the bead tubes with lysate from Step 8 for 12000x g.

13.

Transfer 290µL supernatant without any particle carryover to the wells of the Deep-well plate. This plate becomes the Sample Plate.

14.

Add 135µL Buffer VXL, 540µL Buffer ACB, and 20µL MagAttract Suspension G to each sample in the sample plate. For multiple samples, make a master mix with 10% overage. Invert slowly to mix the master mix; avoid foaming (it can be mixed on a Hula mixer for 2 min). Add 695µL mixture to each sample. Each well including sample lysate should be 985µL.

15.

Select the program IndiMag_Pathogen_KF_Flex_4wash on the instrument.

16.

Start the run, then load the prepared plates into position when prompted by the instrument.

2.3 QUANTIFICATION AND STORAGE

17.

After the running protocol is completed (~35 minutes), immediately remove the elution plate from the instrument and cover the plate or transfer the eluate to the final tube or plate of choice for final storage.

18.

In a 0.6mL microcentrifuge tube, use 3µL total nucleic acid for DNA and RNA concentration measurement using Qubit 4 Fluorometer following manufacturer instructions (Kits needed: Qubit 1X dsDNA HS Assay Kit and Qubit RNA HS Assay Kit) (see Appendix 2 and Appendix 3).

19.

Proceed with sample testing following the REDI-NET SOP SW-4 Swab Testing or store at -20°C for less than 2 weeks (for long-term storage the sample needs to be stored at -80°C following the REDI-NET SOP SW-3 Swab Storage).

3. INSTRUMENT SET UP

20.

Note
NOTE: KingFisher Duo Prime only, if using KingFisher Flex, go to 2.

Confirm 12-tip magnetic heads and 12 well deep-well heat blocks are being used.

21.

Ensure the program IndiMag_Pathogen_KF_Duo_4wash has been downloaded and loaded onto the KingFisher Duo Prime instrument.

3.1 SET UP THE SAMPLE PLATE AND ELUTION STRIP

22.

Set up the Sample Plate according to the table below:

ABCD
Row IDPlate RowReagentVolume per well
Sample rowALysate and lysis buffer985 µL
Wash 1BBuffer AW1700 µL
Wash 2CBuffer AW2700 µL
Wash 3D80% ethanol750 µL
Wash 4E100% ethanol750 µL
Tip CombFTip comb
****GEmpty
****H
23.

Set up the Elution Strip according to the table below:

Note
NOTE: DO NOT use the elution buffer provided by the kit for TNA elution. The ingredients in the elution buffer inhibit the downstream DNA sequencing efficiency.

ABCD
Strip IDRowReagentVolume per well
ElutionANuclease-free water75 µL

3.2 EXTRACTION

24.

Centrifuge the bead tubes with lysate from Step 8 for 12000x g.

25.

Transfer 290µL supernatant without any particle carryover to the wells of the Deep-well Row A. This row becomes the Sample Row.

26.

Add 135µL Buffer VXL, 540µL Buffer ACB, and 20µL MagAttract Suspension G to each sample in the Sample Plate. For multiple samples, make a master mix with 10% overage. Invert slowly to mix the master mix and avoid foaming. Add 695µL mixture to each sample. Each well including sample lysate should be 985µL.

27.

Select the program IndiMag_Pathogen_KF_Duo_4wash on the instrument.

28.

Start the run, then load the prepared plate and Elution Strip into position when prompted by the instrument.

Note
Keep the door open while extraction is in process. The chamber of the KingFisher Duo Prime is small. Closing the door makes the ethanol vapor restrained inside the chamber and increases the ethanol contamination.

3.3 QUANTIFICATION AND STORAGE

29.

After the running protocol is completed (~35 minutes), immediately remove the elution plate from the instrument and cover the plate or transfer the eluate to the final tube or plate of choice for final storage.

30.

In a 0.6mL microcentrifuge tube, use 3µL total nucleic acid for DNA and RNA concentration measurement using Qubit 4 Fluorometer following manufacturer instructions (Kits needed: Qubit 1X dsDNA HS Assay Kit and Qubit RNA HS Assay Kit) (see Appendix 2 and Appendix 3).

31.

Proceed with sample testing following the REDI-NET SOP SW-4 Swab Testing or store at -20°C for less than 2 weeks (for long-term storage the sample needs to be stored at -80°C following the REDI-NET SOP SW-3 Swab Storage).

APPENDIX 2. DNA and RNA Measurement using QUBIT FLUOROMETER 4.0

32.

DNA quantification:

32.1.

According to the volume of sample used, add the 1x HS dsDNA Qubit Assay for a final volume of 200µL (i.e., if using 3µL of sample, add 197µL of 1x HS dsDNA Qubit Assay). Vortex for 5 - 10 seconds, then incubate for 0h 2m 0s at Room temperature

33.

RNA Quantification:

33.1.

In a new microcentrifuge tube/falcon tube (depending on the number of samples processed), prepare a working solution of the Qubit HS RNA Assay:

ABC
ReagentsVolume/sampleVolume for n+1 sample
Qubit RNA HS Assay buffer199 µL…. µL
Qubit RNA HS Assay Dye1 µL…. µL
33.2.

In a new 0.6ml tube, mix 197µL of Qubit HS RNA Assay working solution and 3µL of the sample. Vortex for 5 - 10 seconds, then incubate for 0h 2m 0s at Room temperature before reading.

推荐阅读

Nature Protocols
Protocols IO
Current Protocols
扫码咨询