MicroCT protocols for scanning of embryos and juvenile Hexaplex trunculus

Eva Chatzinikolaou, Kleoniki Keklikoglou

Published: 2021-09-03 DOI: 10.17504/protocols.io.bxw4ppgw

Disclaimer

DISCLAIMER – FOR INFORMATIONAL PURPOSES ONLY; USE AT YOUR OWN RISK

The protocol content here is for informational purposes only and does not constitute legal, medical, clinical, or safety advice, or otherwise; content added to protocols.io is not peer reviewed and may not have undergone a formal approval of any kind. Information presented in this protocol should not substitute for independent professional judgment, advice, diagnosis, or treatment. Any action you take or refrain from taking using or relying upon the information presented here is strictly at your own risk. You agree that neither the Company nor any of the authors, contributors, administrators, or anyone else associated with protocols.io, can be held responsible for your use of the information contained in or linked to this protocol or any of our Sites/Apps and Services.

Abstract

Micro-computed tomography (micro-CT) is a high-resolution 3D-imaging technique which is now increasingly applied in biological studies focusing on taxonomy and functional morphology. The creation of virtual representations of specimens can increase availability of otherwise underexploited and inaccessible samples. This protocol aims to standardise micro-CT scanning procedures for embryos and juveniles of the marine gastropod species Hexaplex trunculus.

Steps

Sample preparation

1.

Removal of embryos of Hexaplex trunculus from their egg capsules.

2.

Anesthetization of embryos and juvenile gastropods with 7% MgCl2.

3.

Sample placement inside a plastic white pipette tip without any scanning medium (in the air).

microCT scanning

4.

Equipment

ValueLabel
SkyScan 1172NAME
micro-computed tomographer (microCT)TYPE
BrukerBRAND
10L01170SKU
Detail detectability: <1 μm
Low contrast resolution (10% MTF): 5 μm
Pixel size at maximum magnification: <0.8 μm

X-ray source: Sealed microfocus X-ray tube, air cooled, >10,000h lifetime; Spot size <5μm @ 4W, 20-100kV, 0-250 μA (10W max)

X-ray detector (camera): 11 Megapixel (4000X2300) 12-bit digital CCD-camera with fibre optic coupling to scintillator

Maximum object size: 50 mm in diameter using offset scan

Radiation safety: < 1 µSv/h at any point on the instrument surface | SPECIFICATIONS |

4.1.

Scanning parameters for embryos and juvenile Hexaplex trunculus

Voltage: 59 kV

Current :167 μA

Filter: none

Pixel size: 2 μm

Camera binning: 1 × 1

Exposure time: 325 ms for embryos and 316 ms for juveniles

Rotation: 360° for embryos and 180° for juveniles

Rotation step: 0.20º for embryos and 0.25° for juveniles

Frame averaging: 3 for embryos and 5 for juveniles

Images reconstruction

5.

Projection images were reconstructed into cross sections using the SkyScan’s NRecon software (Bruker, Kontich, Belgium)

5.1.

Upload projection images

5.2.

Perform X-Y alignment

5.3.

Reconstruction parameters

Smoothing: 2

Ring artifact correction: 20

Beam hardening correction: 59%

Attenuation coefficients: 0 - 1.127 for embryos and 0 - 1 for juveniles

Save as: 16-bit TIFF images