The protocol content here is for informational purposes only and does not constitute legal, medical, clinical, or safety advice, or otherwise; content added to protocols.io is not peer reviewed and may not have undergone a formal approval of any kind. Information presented in this protocol should not substitute for independent professional judgment, advice, diagnosis, or treatment. Any action you take or refrain from taking using or relying upon the information presented here is strictly at your own risk. You agree that neither the Company nor any of the authors, contributors, administrators, or anyone else associated with protocols.io, can be held responsible for your use of the information contained in or linked to this protocol or any of our Sites/Apps and Services.

Abstract

This protocol describes the steps necessary for the preparation and use of screening plates for the rapid detection of terbinafine resistant dermatophytes.

Steps

Terbinafine preparation

1.

Weigh 16.33mg of and add to an erlenmeyer flask containing 5mL of . Mix without heating for 0h 10m 0s to dissolve. The final concentration of this terbinafine preparation is 3200µg/mL . Do not forget to add a magnetic bar. Cover each flask with glass wool and aluminium foil.

Equipment

ValueLabel
Heated magnetic stirrerNAME
AG602TYPE
LMRBRAND
AG602SKU
https://www.labomoderne.comLINK
2.

Autoclave this flask at 121°C for 0h 30m 0s

Equipment

ValueLabel
AutoclaveNAME
TuttnauerBRAND
3850ELSKU
3.

Under hood, transfer 1mL of this preparation into 2 mL eppendorf tubes and store at -80°C for later use.

Equipment

ValueLabel
Biological safety cabinetNAME
AC2-4S8-TUTYPE
EscoBRAND
303110SKU

Medium preparation

4.

Dissolve 30g of in 1L of and let mix on the heated magnetic stirrer for 0h 5m 0s (temperature and mixing speed knob at mid-step)

Equipment

ValueLabel
Heated magnetic stirrerNAME
AG602TYPE
LMRBRAND
AG602SKU
https://www.labomoderne.comLINK
5.

Add 15g of and let dissolve for 0h 10m 0s while mixing and heating

6.

Pour 250mL of this medium into 4 erlenmeyer flasks and label each with final terbinafine concentrations (0.2 µg/mL for the first flask, 0.1 µg/mL for the second, 0.05 µg/mL for the third and 0 µg/mL for the last flask). Given the uncertainty of measurement, it is advisable to work from the highest concentration to the lowest. Do not forget to add a magnetic bar. Cover each flask with glass wool and aluminium foil.

7.

Autoclave these 4 flasks at 121°C for 0h 30m 0s

Equipment

ValueLabel
AutoclaveNAME
TuttnauerBRAND
3850ELSKU

Plates preparation

8.

For the first flask (0.2 µg/mL) : add 15.6µL of a 3200 µg/mL terbinafine preparation (see section number 1). Let mix while heating (temperature and mixing speed knob at mid-step) for 0h 2m 0s . Caution : do not leave for more than 2 minutes otherwise the medium will solidify.

9.

For the second flask (0.1 µg/mL) : add 7.8µL of a 3200 µg/mL terbinafine preparation (see section number 1). Let mix while heating (temperature and mixing speed knob at mid-step) for 0h 2m 0s . Caution : do not leave for more than 2 minutes otherwise the medium will solidify.

10.

For the third flask (0.05 µg/mL) : add 3.9µL of a 3200 µg/mL terbinafine preparation (see section number 1). Let mix while heating (temperature and mixing speed knob at mid-step) for 0h 2m 0s . Caution : do not leave for more than 2 minutes otherwise the medium will solidify.

11.

For the last flask (0 µg/mL) : does not contain terbinafine.

12.

In a 12-well plate (3 rows and 4 columns), the first well of each row does not contain terbinafine. It serves as a growth control. Using an automated pipettor, add 3mL of the medium that does not contains terbinafine (flask number 4) to all the first wells of the plate.

Equipment

ValueLabel
Motorized Pipette FillerNAME
Levo PlusTYPE
DR. LABBRAND
7033100100SKU

.

Equipment

ValueLabel
Microwell plateNAME
Thermo ScientificBRAND
130185SKU
13.

The second wells in each row contain a concentration of 0.05 µg/mL of terbinafine. Using an automated pipettor, add 3mL of the medium that contains 0.05µg/mL of terbinafine (flask number 3) to all the second wells of the plate.

Equipment

ValueLabel
Motorized Pipette FillerNAME
Levo PlusTYPE
DR. LABBRAND
7033100100SKU

Equipment

ValueLabel
Microwell plateNAME
Thermo ScientificBRAND
130185SKU
14.

The third wells in each row contain a concentration of 0.1 µg/mL of terbinafine. Using an automated pipettor, add 3mL of the medium that contains 0.1µg/mL of terbinafine (flask number 2) to all the third wells of the plate.

Equipment

ValueLabel
Microwell plateNAME
Thermo ScientificBRAND
130185SKU

Equipment

ValueLabel
Motorized Pipette FillerNAME
Levo PlusTYPE
DR. LABBRAND
7033100100SKU
15.

The last wells in each row contain a concentration of 0.2 µg/mL of terbinafine. Using an automated pipettor, add 3mL of the medium that contains 0.2µg/mL of terbinafine (flask number 1) to all the last wells of the plate.

Equipment

ValueLabel
Motorized Pipette FillerNAME
Levo PlusTYPE
DR. LABBRAND
7033100100SKU

Equipment

ValueLabel
Microwell plateNAME
Thermo ScientificBRAND
130185SKU
16.

Cover the plates with their lids and let the medium solidify at room temperature. Label the plates with the corresponding terbinafine concentrations for each well and store at 4°C for future use.

Use of the plates

17.

Using a sterile swab, prepare a suspension (sterile water) 0.5 McF of the dermatophyte to be studied.

18.

Transfer 25µL of this suspension to each of the 4 wells of the plate.

19.

Annotate the plate with the number of the studied strains and incubate at 30°C for 96h before reading the result.