U54 SCENT Olink® Target 96
andrew.nixon, Chris Brady
Published: 2023-08-16 DOI: 10.17504/protocols.io.j8nlkookwv5r/v1
Abstract
This protocol outlines the sample preparation and running of the Olink 96 panel
Steps
Preparation
1.
Aliquoting and Day of Preparation
For Aliquoting of EDTA plasma samples: For Aliquoting of EDTA plasma samples :
- Thaw all EDTA plasma samples in a -4°C refrigerator until ready to use.
- Keep Samples in a chilled rack whenever possible.
- Match your parent tubes (Original tubes shipped) with a prelabelled 2mL clarification tube.
- Vortex all samples on a Vortex Genie with a vortex setting of 7-10.
- Transfer samples from the parent tube to the clarification tube.
- In a pre-cooled centrifuge set to -4°C centrifuge all samples for 5 minutes on a setting of 20,000 rcf.
- Carefully remove samples from centrifuge into a chilled rack.
- Transfer liquid from the clarification tube to a prelabelled aliquot tube in desired amounts.
- Using Liquid Nitrogen, freeze back the tubes and place in a -80°C freezer until ready to use.
For EDTA samples on day of use: For EDTA samples on day of use se:
-
Thaw all EDTA plasma samples in a -4°C refrigerator until ready to use.
-
Keep Samples in a chilled rack whenever possible.
-
Vortex all samples on a Vortex Genie with a vortex setting of 7-10.
-
In a pre-cooled centrifuge set to -4°C centrifuge all samples for 5 minutes on a setting of 20,000 rcf.
-
Return the samples to a chilled rack.
-
Your samples are ready for use.
Running Olink Panel
2.
Once samples are prepared use the following links to access the protocols: