Size Exclusion Chromatography with Multiangle Light Scattering (SEC-MALS)

Xuefeng Ren, Minghao Chen

Published: 2023-10-17 DOI: 10.17504/protocols.io.j8nlkom4xv5r/v1

Abstract

SEC-MALS

Steps

1.

Concentrate the purified protein sample to 4-5 mg/ml

2.

Instruments:

Agilent 1200 HPLC system (Agilent Technologies, Santa Clara, CA)

Wyatt DAWN HELEOS-II MALS instrument

Wyatt Optilab rEX differential refractometer (Wyatt, Santa Barbara, CA)

WTC-050S5 size-exclusion column (Wyatt)

40 ul sample loop

3.

Equilibrate the system and column with buffer containing:

25 mM HEPES pH 7.5, 200 mM NaCl, 1 mM MgCl2, 2 mM TCEP

4.

Turn on the laser 1 hour before the experiment.

Confirm the readout is stable

5.

Run 5 mg/ml BSA to calibrate the system

6.

Inject the sample and run (takes ~1 hour/sample)

7.

Analyze the results by ASTRA V software (Wyatt).

MALS signals, combined with the protein concentration determined by refractive index, are used to calculate the molecular mass of the complex.