Electrophysiology from cervical vagus nerve and great auricular nerve in swine
Kip Ludwig, Nishant Verma
Abstract
This protocol was used to collect data in preparation for publication (Characterization of Electrodes to Record Neural Signals in the Periphery). Additionally, the dataset is publicly available on Pennsieve.
In this study, we characterized cuff electrodes, longitudinal intrafascicular electrodes (LIFEs), an intrafascicular electrode used in pre-clinical studies (Yoshida and Stein, 1999, DOI: 10.1109/10.740885; Nicolai et al., 2020, DOI: 10.1088/1741-2552/ab9db8), and microneurography electrodes, a microelectrode used clinically to measure muscle sympathetic nerve activity (MSNA), in their ability to measure neural activity in the peripheral nervous system. The cervical vagus, a major autonomic nerve innervating organs in the thorax and abdomen as well as muscles in the throat, and the great auricular nerve, a sensory nerve innervating the ear and periauricular region, were used as model nerves.
Before start
Consult your institution's Institutional Animal Care and Use Committee (IACUC) before attempting any of the following in a live animal.
Steps
Institutional Animal Care and Use Committee Approval
Seek approval from your local IACUC.
Anesthesia
Deliver intramuscular injection of Telazol (6 mg/kg) and Xylazine (2 mg/kg) to induce sedation.
Ventilate the animal and maintain the surgical plane with inhaled Isoflurane (~1-2%) and intravenous Fentanyl (12-30 mcg/kg/hr) administered with lactated Ringer’s solution (LRS) (ICU medical, IM-4389).
Cervical vagus nerve preparation
With the subject in a supine position, use a midline approach to access the left carotid sheath. The carotid artery is mobilized and carefully retracted to minimize obstruction to blood flow. The cervical vagus nerve is exposed for a length of 9-12 cm.
Instrument the vagus nerve with a bipolar stimulation electrode caudal to the superior laryngeal branching. Three replicates of the Longitudinal intrafascicular electrode (LIFE) and microneurography electrodes and a cuff electrode with three recording contacts are instrumented on the nerve caudal to the stimulation electrode. A separation of >4 cm is kept between the stimulation electrode and the closest recording electrode.
A reference LIFE electrode and microneurography electrode are inserted in superficial fat. The reference electrode site is selected at a distance approximately equidistant from the stimulation electrode to the recording electrodes to match the representation of the stimulation artifact.
Evoked compound action potentials (ECAPs) are recorded by delivering 750 biphasic stimulation pulses at 25 Hz and 200 us pulse width with randomized stimulation amplitude between 0 and 10 mA. Time locked recordings are made through the recording electrodes.
Great auricular nerve preparation
The great auricular nerve (GAN) is accessed by the following approach: The skin and subcutaneous fat are incised from the medial posterior margin of the ramus. A small notch is palpated at this point and indicates the approximate level of the stylomastoid foramen. Here the facial nerve exits and divided into its various branches.
The incision continues dorsal following a line posterior to the temporomandibular articulation and up to the medial base of the ear just inferior to the medial crus of the helix to follow the lateral vein through the base of the ear as a landmark for finding the sensory input for the exterior skin of the auricle. This incision exposes the superficial musculoaponeurotic system (SMAS). The SMAS layer is divided along the posterior margin of the ramus and posterior margin of the temporomandibular articulation process to expose the subfascial level adipose tissue. This underlying adipose tissue along the posterior margin of the temporomandibular articulation is the location where the GAN and branches of the facial nerve are found.
To determine which of the exposed branches were motor or sensory in origin, electrically stimulate each of the nerve branches to look for a motor response.
When the sensory branch of the posterior auricular nerve is identified by no motor response it is instrumented with a stimulation cuff and recording electrodes (two microelectrodes and a two-contact cuff or two LIFE electrodes) and verified by recording sensory evoked responses from the skin at the base of the auricle. Reference electrodes are inserted similarly as described above in the vagus preparation.
Sensory-evoked naturally occurring neural activity are recorded by stroking the region of the ear innervated by the GAN with a brush.
Transcutaneous electrical nerve stimulation (TENS) electrodes cut to 2 x 2 cm in size are applied to the same region of sensory innervation of the GAN as confirmed by recording of sensory evoked potentials and at the base of the ear through which the main trunk of the GAN courses.
Non-invasive ECAPs are recorded by delivering stimulation parameters identical to those reported in the cervical vagus section. Invasive ECAPs were recorded by delivering 250 biphasic stimulation pulses at 25 Hz and 200 us pulse width with randomized stimulation amplitude between 0 and 3-10 mA (determined based on each subject’s motor threshold). Time locked recordings are made through the recording electrodes.
Electrophysiology system
A Tucker-Davis Technologies (TDT) electrophysiology system is used for stimulation and recording. The front-end and digitization is done on the battery-powered Subject Interface (SIM). Data is collected at 25 kHz.
High impedance (microneurography electrode) and low impedance (LIFE and cuff) are recorded into two separate recording cards with an active and passive head stage respectively.
Stimulation is delivered in bipolar mode with two floating current sources.
Electrophysiology data analysis
A custom-built and publicly available Python package, PyeCAP (https://github.com/ludwig-lab/pyeCAP), is used for offline analysis of electrophysiology, ECAP, and physiology data.
Filtering. Electrophysiology data for ECAP analysis was filtered with a high pass 1st order Butterworth filter with a corner frequency at 100 Hz and a low pass Gaussian filter with a corner frequency at 3 kHz. An additional 60 Hz band stop finite impulse response (FIR) filter constructed with a Hamming window is used on the electrophysiology data for analysis of naturally occurring activity. All filtering is performed on the time series data in both the forward and backward direction to eliminate group delays caused by filtering.
Detecting Authentic ECAPs. ECAPs were plotted by averaging (point by point median) the stimulation evoked response across all available consecutive pulses in a particular stimulation train. ECAPs are detected by windowing the trace into time windows according to the distances between recording and stimulation electrode and published conduction speeds by fiber type (Manzano et al., 2008, 10.1590/s0004-282x2008000100033).
ECAP authenticity is confirmed by several methods. Firstly, conduction speed delay across recording electrodes are verified to be in the expected nerve fiber conduction speed range (Manzano et al., 2008, 10.1590/s0004-282x2008000100033)
Quantifying ECAP strength. Aβ- and B-fiber ECAPs are quantified to investigate their strength. Aβ fibers are used as a measure of fast conducting fibers and B-fibers are used as a measure of slower conducting myelinated fibers.
Aβ- and B-fiber ECAP time windows are calculated for each stimulation train based on distance from the recording electrode to the stimulation electrode and published values of conduction speed (Manzano et al., 2008, 10.1590/s0004-282x2008000100033). The window are then manually narrowed, without removing the ECAP signal of interest, to be of equal duration across recording channels.
Spike detection. Spike detection is done using a thresholding method. Standard deviation (SD) from t=1-6s of the specific recording on the specific channel is calculated and the threshold is set at six times this SD plus the mean, to account for baseline offsets, from t=1-6s of the same recording.